Messung einzelner Mikrolitertröpfchen

Anwendungsbeispiel

Introduction

This application note shows how the Pulsed Flow Sensor (PFS) can be used in combination with a micro valve (B-drop mini) to measure single droplet volumes from 0.2 to 16 µl.

Used Material

  • Pulsed Flow Sensor
  • PFS controller with User Interface
  • B-drop mini
  • Pressure supply 0.5 bars
  • Water container 300 ml
  • Silicone Tubing
droplet measurement

Experiment setup

The PFS was connected to a B-drop mini in a vertical setup. The tube between sensor and valve was kept as short as possible to prevent damping.

The system was then pressurized to 0.5 bars (without pressure control) and the measurement mode was set to “Droplet Mode”. Then the valve was triggered with different opening times using the following waveform.

[:de]peak hold graph[:]

Messungen

Für 5 verschiedene Öffnungszeiten wurden die Tropfenvolumina mit dem Pulsed Flow Sensor gemessen. Die Messdiagramme zeigen, dass bei der kürzesten Öffnungszeit (2 ms) das Ventil bereits schließt, bevor ein konstantes Flussplateau erreicht werden kann. Dennoch kann eine gute Wiederholbarkeit erreicht werden. Längere Öffnungszeiten führen zu einem geringeren Einfluss des Öffnungs- und Schließvorgangs des Mikroventils und zeigen daher eine bessere Wiederholbarkeit.

[:de]Time-Flow graph measuring single microliter droplets[:]

Gemessene Daten

[:de]Opening time of the valve chart[:]