The HFS is a sensor with a hygienic design that is easy to flush. It is based on the highly dynamic differential pressure technology of the PFS and can measure up to 10 L/min. The combination of the HFS with a dynamic valve is ideal for dosing small quantities at high frequency (up to 50 Hz).
The HFS measures with a sampling frequency of up to 20 kHz and can resolve volumes from 0.1 ml to L with high accuracy.
Features:
- Dynamic measurement with response times < 1 millisecond
- Hygienic design
- Volume flow measurement up to 10 L/min
- Exact volume integrals from 0.1 ml
- For water, ethanol and other low-viscosity media (up to 50 mPas)
- Not susceptible to gas bubbles
- Integrated temperature and pressure measurement
- Suitable for the food sector
- Sterilization by hot flush up to 70°C and 3 bar and with ethanol
Technical Data
Ordering Nr. | HFS-E10-3.5 (0-4L/min)/ HFS-E10-5.2 (0-12L/min) |
Connectors | DIN 11864-1 form A DN15 |
Sensing modes | Flow, volume integral, pressure, temperature |
Materials in contact with media | 1.4404 (316L), PEEK(FDA), FKM (FDA,USP Class VI,EC 1935) |
Accuracy of flow rate | better than ± 2 % FS |
Media calibration | Programmed viscosity and density |
Communication | RS485 |
Supply Voltage | 24V |
Parameters | Symbol | Min. | Max. | Unit |
Calibrated flow rate | V’ | 0.4
1.2 |
4.0
12.0 |
L/min
L/min |
Pressure range | pw | 0.5 | 4.0 | bar (absolute) |
Burst pressure | pb | 10 | bar | |
Working temperature range | Tw | 0 | 50 | °C |
Cleaning temperature range | Tc | 0 | 70 | °C |